STM Calibration
STM Cal. tab allow user to adjust calibration data for tunneling mode
of SPM operation.
- Max It, nA - maximum tunneling current. This value depends upon a sample
table that is used (with hi or low sensitivity). This value can be
calculated fom following equation:
,where
is sample table resistance.
- Max Ut, V - Maximum voltage on the tunneling gap.
- Inverse feedback - sign of a feedback.
With positive feedback incriase in error signal leads to increase in
feedback signal, with negative feedback - increase in error signal leads
to decrease in feedback signal. In STM mode positive feedback is used.
- Inverse It - Error signal sign. In STM head inverting operational amplifier is
installed. So the error signal come to ADC in inverted form.
- Feedback channel - channel of 1-st ADC, from which feedback signal is
taken. At the present moment 0 channel is used for error (deflection) signal, and
1-st channel used for RMS signal (which is used as feedback signal in
Resonance AFM mode)