Sample landing preferences
Among numbers on a Values tab user can change following ones:
- Landing It - tunneling current value (in % from that was set as
bearing value for scanning) that if reached stop the tip to sample landing process.
- Landing delay - number of delay cycles corresponding to singe motor step during
ordinary (started by client) landing process. Usually 5000 cycles are corresponded to
one second. The less the landing delay is, the faster a sample moves. However too
little landing delay may cause instability in stepper motor work.
- Landing length - maximum number of stepper motor steps. If reached, it stops landing
and user is asked if he wants to continue landing.
- Landing step - precision of sample positioning, observed by client.
- Draw back on every step - landing mode in which it check if it can reach
the surface. During landing process before every step piezomanipulator fully
retract the sample. After the step piezomanipulator returns to initial position.
If during the last operation feedback become active then landing consider
to be finished.
- Draw delay - delay (in DSP cycles) between steps during landing/retracting by
piezomanipulator. Value 0 corresponds to absence of this delay.
- Draw stepping - value of the piezomanipulator step during landing/retracting by
piezomanipulator.